Loading…

The measurement of right ventricular ejection fraction by thermodilution : a comparison of values obtained using differing injectate ports

Thermodilution measurements of cardiac output and right ventricular ejection fraction were obtained using a rapid response pulmonary artery catheter. Values were compared when injectate was administered via either a cannula within the right internal jugular vein or the dedicated right atrial port of...

Full description

Saved in:
Bibliographic Details
Published in:Anaesthesia 1993-04, Vol.48 (4), p.312-314
Main Authors: COCKROFT, S, WITHINGTON, P. S
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Thermodilution measurements of cardiac output and right ventricular ejection fraction were obtained using a rapid response pulmonary artery catheter. Values were compared when injectate was administered via either a cannula within the right internal jugular vein or the dedicated right atrial port of the pulmonary artery catheter. Mean (SD) bias for cardiac output and right ventricular ejection fraction measurements were 0.08 (0.32) l.min-1 and 2.6 (6.6)% respectively. We therefore conclude that both injectate techniques will provide similar values for cardiac output but dissimilar values for right ventricular ejection fraction measurement.
ISSN:0003-2409
1365-2044
DOI:10.1111/j.1365-2044.1993.tb06950.x