Loading…
Noninvasive measurement of cervical tri-planar motion in normal subjects
An electronic digital inclinometer was used to obtain normative data on cervical mobility from 58 normal subjects. They consisted of 28 males (age range, 17-61) and 30 females (age range, 19-62). The measurements were taken from occiput to T1 in flexion-extension, lateral bend, and axial rotation, i...
Saved in:
Published in: | Spine (Philadelphia, Pa. 1976) Pa. 1976), 1993-11, Vol.18 (15), p.2191-2195 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An electronic digital inclinometer was used to obtain normative data on cervical mobility from 58 normal subjects. They consisted of 28 males (age range, 17-61) and 30 females (age range, 19-62). The measurements were taken from occiput to T1 in flexion-extension, lateral bend, and axial rotation, isolating the regional motion by measuring simultaneously the upper and lower margins of the eight mobile segments. A comparison of radiographs and inclinometer measures showed excellent correlations (r < 0.9997, P < 0.05). Normative data are presented. Gender-specific regional mobility differences occurred only in cervical extension, whereas no consistent age-related differences were found. Stiffer individuals could be found in either gender or any age group, but rigidity in one plane did not necessarily predict rigidity in another. |
---|---|
ISSN: | 0362-2436 1528-1159 |
DOI: | 10.1097/00007632-199311000-00007 |