Paradoxical Induction of Endless Loop Tachycardia by Magnet Application Over a DDD Pacemaker

It is well known that removal of a testing magnet from a DDD pulse generator may cause endless loop tachycardia in patients with retrograde ventriculoatrial conduction; application of the magnet then terminates the tachycardia. We have observed the opposite response to the magnet and in this report...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 1986-07, Vol.9 (4), p.503-510
Main Authors: BAROLD, S. SERGE, FALKOFF, MICHAEL D., ONG, LING S., HEINLE, ROBERT A.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:It is well known that removal of a testing magnet from a DDD pulse generator may cause endless loop tachycardia in patients with retrograde ventriculoatrial conduction; application of the magnet then terminates the tachycardia. We have observed the opposite response to the magnet and in this report we describe the paradoxical induction of endless loop tachycardia by magnet application over a DDD pulse generator and its persistence despite repeated removal and reapplication of the magnet. This unusual behavior occurred only in the “magnet off” function and is due to magnet‐induced signals sensed by the atrial channel circuitry.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1986.tb06606.x