Late complication of a Greenfield filter associating caudal migration and perforation of the abdominal aorta by a ruptured strut

We report the case of a 67-year-old woman who was admitted for surgical removal of a Greenfield filter that had been inserted 7 years before because of recurrent deep vein thrombosis associated with pulmonary embolism. This complication appeared on a plain abdominal radiogram that showed a 7 cm dist...

Full description

Saved in:
Bibliographic Details
Published in:Journal of vascular surgery 1995-08, Vol.22 (2), p.182-187
Main Authors: Dabbagh, Ahmad, Chakfé, Nabil, Kretz, Jean-Georges, Demri, Boualem, Nicolini, Philippe, Fuentes, Claudio, Mettauer, Bertrand, Epailly, Eric, Muster, Dominique, Eisenmann, Bernard
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report the case of a 67-year-old woman who was admitted for surgical removal of a Greenfield filter that had been inserted 7 years before because of recurrent deep vein thrombosis associated with pulmonary embolism. This complication appeared on a plain abdominal radiogram that showed a 7 cm distal migration of the filter, a 30-degree angulation, and rupture of a strut at the level of the hub. Computed tomography, aortography, and ascending cavography demonstrated that the inferior vena cava was perforated by the struts and that the ruptured strut had penetrated the infrarenal aorta. As demonstrated by scanning electron microscopy, the fracture was due to a structural defect of the strut at its insertion point within the hub, with no sign of corrosion. Energy-dispersive radiography analysis failed to demonstrate impurity in the metal composition. (J V ASC S URG 1995;22:182-7.)