Loading…
Low inductance transverse gradient system of restricted length
A new transverse gradient coil assembly of restricted length is presented. The coil is symmetric, has the advantage of simplicity, generates a remarkably large volume of uniform transverse gradient field, features very low inductance, and can therefore be suitable for applications requiring fast swi...
Saved in:
Published in: | Magnetic resonance imaging 1995, Vol.13 (4), p.607-613 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new transverse gradient coil assembly of restricted length is presented. The coil is symmetric, has the advantage of simplicity, generates a remarkably large volume of uniform transverse gradient field, features very low inductance, and can therefore be suitable for applications requiring fast switched gradients. A prototype coil has been constructed to check computer simulations and to compare measured parameters of the system with those expected. Coils of this type may be used for MRI of the human head, and of animals, and for NMR microimaging. |
---|---|
ISSN: | 0730-725X 1873-5894 |
DOI: | 10.1016/0730-725X(95)00002-X |