Loading…

Olbert dilatation catheter system: complication due to inexpert handling

A complication of inexpert handling of the Olbert catheter system is presented. The balloon was inflated while the clamp at the proximal end of the catheter was closed. This resulted in kinking of the distal end of the catheter within the balloon. Difficulty in catheter removal resulted but was over...

Full description

Saved in:
Bibliographic Details
Published in:Cardiovascular and interventional radiology 1987-01, Vol.10 (3), p.168-170
Main Authors: Grabenwöger, F, Dock, W, Pinterits, F
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A complication of inexpert handling of the Olbert catheter system is presented. The balloon was inflated while the clamp at the proximal end of the catheter was closed. This resulted in kinking of the distal end of the catheter within the balloon. Difficulty in catheter removal resulted but was overcome by an improvised coaxial sheath technique.