Blind placement of plastic left double-lumen tubes

A prospective analysis of placement of left-sided plastic double-lumen tubes in 100 patients is presented. Intubation of the left bronchus was successfully accomplished using only auscultation and clinical signs ("blind" placement) in 91 patients. Double-lumen tubes were positioned in less...

Full description

Saved in:
Bibliographic Details
Published in:Anaesthesia and intensive care 1995-10, Vol.23 (5), p.583-586
Main Authors: BRODSKY, J. B, MACARIO, A, CANNON, W. B, MARK, J. B. D
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A prospective analysis of placement of left-sided plastic double-lumen tubes in 100 patients is presented. Intubation of the left bronchus was successfully accomplished using only auscultation and clinical signs ("blind" placement) in 91 patients. Double-lumen tubes were positioned in less than five minutes in 84 patients. The most common problem encountered (30%) was initial intubation of the right main bronchus. Seven of these patients required bronchoscopic assistance to guide the tube into the left bronchus. There were four minor intraoperative complications due to DLT malposition that were recognized and corrected by withdrawing the tube slightly back in the bronchus. The plastic double-lumen tubes functioned properly during the procedure in all 100 patients.