Environmental Electromagnetic Interference from Electronic Article Surveillance Devices: Interactions with an ICD

A patient with an implantable cardioverter defibrillator (ICD) experienced an inappropriate firing while in close proximity to an electronic article surveillance (EAS) device. Testing revealed the ICD was able to detect high frequency “noise” when close to the EAS device. ICD patients may need couns...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 1995-12, Vol.18 (12), p.2229-2230
Main Author: McIVOR, MICHAEL E.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A patient with an implantable cardioverter defibrillator (ICD) experienced an inappropriate firing while in close proximity to an electronic article surveillance (EAS) device. Testing revealed the ICD was able to detect high frequency “noise” when close to the EAS device. ICD patients may need counseling to avoid close contact with such devices.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1995.tb04654.x