The accuracy of the flowrate in flush-devices of disposable pressure transducers
Arterial and venous pressure is commonly measured using fluid filled catheters. To avoid obstruction they are continuously spilled by a flush-device. The accuracy of the flowrate has not been investigated previously. The accuracy of 5 different flush-devices available in Switzerland was checked for...
Saved in:
Published in: | International journal of clinical monitoring and computing 1995-01, Vol.12 (4), p.219-224 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Arterial and venous pressure is commonly measured using fluid filled catheters. To avoid obstruction they are continuously spilled by a flush-device. The accuracy of the flowrate has not been investigated previously.
The accuracy of 5 different flush-devices available in Switzerland was checked for flowrate when factory new, after a single sterilization with ethylen-oxide, in a long-term test over 96 hours, after repeated handling of the integrated bypass and under application of a pulsing counter-pressure.
Flow is linearly related to differential-pressure and is constant over time. The flow of each flush-device at 200 mmHg differential-pressure was below the indicated 3 ml per hour (1.69 - 2.49 ml/h). Sterilization in two types produced a significant but not relevant difference in flowrate. Longtime-use, bypass actuation and pulsing pressure did not alter the flowrate significantly (p |
---|---|