Percutaneous endoluminal stent-graft repair of an old traumatic femoral arteriovenous fistula
A stent-graft was custom made to close a high-flow traumatic arteriovenous fistula of the left superficial femoral artery, present for 30 years, in a 60-year-old man with congestive heart failure and ischemic ulceration in the left foot. A balloon expandable Palmaz stent (P394; 2.5 mm X 3.9 cm) was...
Saved in:
Published in: | Cardiovascular and interventional radiology 1996-03, Vol.19 (2), p.120-122 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A stent-graft was custom made to close a high-flow traumatic arteriovenous fistula of the left superficial femoral artery, present for 30 years, in a 60-year-old man with congestive heart failure and ischemic ulceration in the left foot. A balloon expandable Palmaz stent (P394; 2.5 mm X 3.9 cm) was covered with a polytetrafluoroethylene (PTFE) graft and was inserted percutaneously through an 11 Fr vascular sheath. Follow-up Doppler ultrasound at 6 months demonstrated occlusion of the arteriovenous fistula, patency of the artery, and luminal integrity of the artery and vein. |
---|---|
ISSN: | 0174-1551 1432-086X |
DOI: | 10.1007/BF02563907 |