Flow cytometric analysis of 68 monoclonal anti-D reagents
68 monoclonal IgG anti-D were assessed for quantitative and qualitative binding to D variant red cells using a flow cytometric indirect immunofluorescence test. One antibody failed to bind to normal RhD pos. cells, but reacted weakly with D cat VI and D Cat VII cells. Quantitatively, binding varied...
Saved in:
Published in: | Transfusion clinique et biologique (Paris) 1996, Vol.3 (6), p.415-417 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | 68 monoclonal IgG anti-D were assessed for quantitative and qualitative binding to D variant red cells using a flow cytometric indirect immunofluorescence test. One antibody failed to bind to normal RhD pos. cells, but reacted weakly with D cat VI and D Cat VII cells. Quantitatively, binding varied approximately twenty-fold between different MAbs and different D variants. D cat III cells gave the lowest level of binding when compared with the D pos. controls. Qualitatively, R
1
VIr, R
2
VIr, R
1
VIIr and DFR samples failed to react with various MAbs regardless of the levels of binding of these MAbs to the D pos. controls, thus supporting the idea of missing epitopes in such variants. |
---|---|