Loading…

Silicon Absolute X-Ray Detectors

The responsivity of silicon photodiodes having no loss in the entrance window, measured using synchrotron radiation in the 1.75 to 60 keV range, was compared to the responsivity calculated using the silicon thickness measured using near-infrared light. The measured and calculated responsivities agre...

Full description

Saved in:
Bibliographic Details
Published in:Proceedings of the 10th International Conference 2009-10, Vol.1234, p.842-845
Main Authors: Seely, John F, Korde, Raj, Sprunck, Jacob, Medjoubi, Kadda, Hustache, Stephanie
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The responsivity of silicon photodiodes having no loss in the entrance window, measured using synchrotron radiation in the 1.75 to 60 keV range, was compared to the responsivity calculated using the silicon thickness measured using near-infrared light. The measured and calculated responsivities agree with an average difference of 1.3%. This enables their use as absolute x-ray detectors.
ISSN:0094-243X