Ventricular Output Failure in a DDD Permanent Pacemaker Associated with Increased Atrial Output

Previous reports have described the occurrence of ventricular output failure in a permanent DDD pacemaker system related to an increase in the atrial output in the presence of low atrial lead impedance (Medtronic Synergyst/Synergyst II). This phenomenon is seen exclusively following atrial paced eve...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 1997-03, Vol.20 (3), p.739-742
Main Authors: SANTUCCI, PETER A., MITRA, RAMAN L.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Previous reports have described the occurrence of ventricular output failure in a permanent DDD pacemaker system related to an increase in the atrial output in the presence of low atrial lead impedance (Medtronic Synergyst/Synergyst II). This phenomenon is seen exclusively following atrial paced events, and may potentially lead to significant bradyarrhythmia or ventricular asystole in a pacemaker dependent patient. We describe tbe occurrence of analogous behavior in a Medtronic Symbios 7006 generator.