Ventricular Output Failure in a DDD Permanent Pacemaker Associated with Increased Atrial Output
Previous reports have described the occurrence of ventricular output failure in a permanent DDD pacemaker system related to an increase in the atrial output in the presence of low atrial lead impedance (Medtronic Synergyst/Synergyst II). This phenomenon is seen exclusively following atrial paced eve...
Saved in:
Published in: | Pacing and clinical electrophysiology 1997-03, Vol.20 (3), p.739-742 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Previous reports have described the occurrence of ventricular output failure in a permanent DDD pacemaker system related to an increase in the atrial output in the presence of low atrial lead impedance (Medtronic Synergyst/Synergyst II). This phenomenon is seen exclusively following atrial paced events, and may potentially lead to significant bradyarrhythmia or ventricular asystole in a pacemaker dependent patient. We describe tbe occurrence of analogous behavior in a Medtronic Symbios 7006 generator. |
---|---|