Limitations of the algorithm to reconfirm ventricular fibrillation in a third-generation implantable cardioverter defibrillator
We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia...
Saved in:
Published in: | Pacing and clinical electrophysiology 1997-04, Vol.20 (4 Pt 1), p.993-994 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia escape interval is required to inhibit the shock. |
---|---|