Loading…

Limitations of the algorithm to reconfirm ventricular fibrillation in a third-generation implantable cardioverter defibrillator

We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 1997-04, Vol.20 (4 Pt 1), p.993-994
Main Authors: Mattke, S, Kramm, B
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia escape interval is required to inhibit the shock.
ISSN:0147-8389
DOI:10.1111/j.1540-8159.1997.tb05507.x