Stimulus artefact in somatosensory evoked potential measurement

When an electrical stimulus is used to evoke action potentials in peripheral nerves or the spinal cord, the stimulus causes an artefact which may interfere with measurement of the evoked potentials. This artefact, unlike all other sources of noise in the measurement, cannot be reduced by ensemble av...

Full description

Saved in:
Bibliographic Details
Published in:Medical & biological engineering & computing 1997-05, Vol.35 (3), p.211-215
Main Authors: SCOTT, R. N, MCLEAN, L, PARKER, P. A
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:When an electrical stimulus is used to evoke action potentials in peripheral nerves or the spinal cord, the stimulus causes an artefact which may interfere with measurement of the evoked potentials. This artefact, unlike all other sources of noise in the measurement, cannot be reduced by ensemble averaging. Confusion about the origin and transmission of stimulus artefact has led to considerable frustration in spinal somatosensory evoked potential (SSEP) measurements. The three mechanisms by which stimulus artefact is coupled into the measuring system are identified, and means for their reduction are discussed.