Loading…
Comparison of Computer Simulated and Phantom Measured Phase Variance in the Study of Trabecular Bone
A new method using magnetic resonance phase images for the assessment of trabecular bone structure has recently been proposed. To test this method, a mathematical model is developed which calculates the phase distribution in gradient echo acquired phase images of a structure of Pyrex glass rods imme...
Saved in:
Published in: | Magnetic resonance imaging 1998, Vol.16 (1), p.29-36 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new method using magnetic resonance phase images for the assessment of trabecular bone structure has recently been proposed. To test this method, a mathematical model is developed which calculates the phase distribution in gradient echo acquired phase images of a structure of Pyrex glass rods immersed in a copper sulfate solution. Several experiments were performed using a phantom built in the same way as the structure used in the mathematical model. The results from the model are compared with those from the phantom tests, and the influence of resolution and bone area fraction on the phase dispersion is studied. The good correlation between theoretical and experimental results shows that phase variance increases with increasing resolution and bone density. However, the dependence of variance on bone density is less prominent for large pixel sizes. |
---|---|
ISSN: | 0730-725X 1873-5894 |
DOI: | 10.1016/S0730-725X(97)00218-X |