Loading…

Intraoperative Recordings of Monophasic Action Potentials with Chronically Implantable Pacemaker Leads

Since the development of fracially coaled Iridium electrodes recordings of monophasic action potentials are possible. Intraoperative recordings of MAP from 15 pacemaker implantations were done after positioning the chronically implantabie pacemaker leads (Biotronik, Berlin, Germany) in the right ven...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 1998-01, Vol.21 (1), p.235-238
Main Authors: ZRENNER, BERNHARD, MÜSSIG, DIRK, SCHREIECK, JÜRGEN, WEYERBROCK, SONJA, SCHNEIDER, MICHAEL, SCHALDACH, MAX, SCHÖMIG, ALBERT, SCHMITT, CLAUS
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Since the development of fracially coaled Iridium electrodes recordings of monophasic action potentials are possible. Intraoperative recordings of MAP from 15 pacemaker implantations were done after positioning the chronically implantabie pacemaker leads (Biotronik, Berlin, Germany) in the right ventricle by using five screw‐in electrodes and ten anchor electrodes. Intraoperative recordings of MAP are possible with all implanted pacemaker leads. The recordings of typical MAP signals were always accompanied with stable electrode positions and good sensing and pacing characteristics.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1998.tb01095.x