Loading…
Fully automated assay for total homocysteine, cysteine, cysteinylglycine, glutathione, cysteamine, and 2-mercaptopropionylglycine in plasma and urine
We describe a 6-min HPLC method to measure the total concentrations of the most important thiols in plasma and urine--cysteine, homocysteine, cysteinylglycine, and glutathione--as well as the concentrations in plasma and urine, respectively, of cysteamine and 2-mercaptopropionylglycine, two compound...
Saved in:
Published in: | Clinical chemistry (Baltimore, Md.) Md.), 1998-04, Vol.44 (4), p.825-832 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We describe a 6-min HPLC method to measure the total concentrations of the most important thiols in plasma and urine--cysteine, homocysteine, cysteinylglycine, and glutathione--as well as the concentrations in plasma and urine, respectively, of cysteamine and 2-mercaptopropionylglycine, two compounds used to treat disorders of cysteine metabolism. Precolumn derivatization with bromobimane and reversed-phase HPLC were performed automatically by a sample processor. Throughput was up to 100 samples in 24 h. The within-run CV ranged from 0.9% to 3.4% and the between-run CV ranged from 1.5% to 6.1%. Analytical recovery was 97-107%, with little difference between plasma and urine samples. The detection limit was approximately 50 nmol/L for all the analytes studied. Thiol concentrations were determined in the plasma of 206 healthy donors and in the urine of 318 healthy donors distributed for age and sex. Mean values of plasma cysteine and homocysteine were significantly lower in infants (ages, |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1093/clinchem/44.4.825 |