Comparison of newborn hearing screening by transient otoacoustic emissions and auditory brainstem response using ALGO-2
The aim of this study is to compare pass rates for two different hearing screening methods in well newborns as a function of age. A previous study by this group compared click evoked otoacoustic emissions (EOAE) and automated auditory brainstem response (ABR) using the ALGO-1® infant hearing screene...
Saved in:
Published in: | International journal of pediatric otorhinolaryngology 1998-05, Vol.43 (3), p.207-211 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The aim of this study is to compare pass rates for two different hearing screening methods in well newborns as a function of age. A previous study by this group compared click evoked otoacoustic emissions (EOAE) and automated auditory brainstem response (ABR) using the ALGO-1® infant hearing screener (Natus Medical, Foster City, CA). Since that study, a new generation automated ABR screener, the ALGO-2®, has been developed. In this study, 232 ears in 116 healthy newborn infants aged 5–48 h were tested using the ALGO-2 screener and EOAE. Overall, 92% of ears passed the ABR, while 57% passed the EOAE screen. The ABR pass rate was higher than in the previous study, where 88.5% of ears passed the ABR screen using ALGO-1. The EOAE pass rate in the present study was lower than in the previous study, in which 79% passed the EOAE screen. Pass rates for both EOAE and ABR improved significantly with increasing infant age. There was no significant difference in the test time required for ALGO-2 (5.7 min) compared with EOAE (5.2 min). The results are compared with earlier studies and implications for universal hearing screening are discussed. |
---|---|
ISSN: | 0165-5876 1872-8464 |
DOI: | 10.1016/S0165-5876(98)00007-X |