Cementing Techniques for Hip Resurfacing Arthroplasty: In Vitro Study of Pressure and Temperature

Abstract Cementing irregularities have been associated with femoral failures of resurfacing arthroplasties in retrieval studies. We used an in vitro model to measure pressure, temperatures, and cement penetration as a function of 6 different cementing techniques. Filling the component with cement ca...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty 2011, Vol.26 (1), p.144-151
Main Authors: Bitsch, Rudi G., MD, Loidolt, Travis, BA, Heisel, Christian, MD, Schmalzried, Thomas P., MD
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract Cementing irregularities have been associated with femoral failures of resurfacing arthroplasties in retrieval studies. We used an in vitro model to measure pressure, temperatures, and cement penetration as a function of 6 different cementing techniques. Filling the component with cement can lead to overpenetration or increase the resistance to component seating with resultant polar cement mass. Both conditions result in high and long-lasting cement pressures, cement defects, as well as peak temperatures higher than 50°C. Manual application of cement provides complete penetration of the available fixation area with the lowest cement pressures, the smallest total cement mass, and a peak temperature of 36.0°C ± 4.1°C. Application of the principles elucidated by this study may reduce the risk of cement overpenetration and incomplete seating.
ISSN:0883-5403