False-negative test results in the Slidex Staph Plus (bioMérieux) agglutination test are mainly caused by spa-type t001 and t001-related strains

The relative sensitivity of commercial agglutination kits for fast identification of S. aureus is usually given to be about 98%. This reported sensitivity has sometimes been questioned. In this study, three collections of molecularly defined, single-copy strains of S. aureus were used to compare the...

Full description

Saved in:
Bibliographic Details
Published in:European journal of clinical microbiology & infectious diseases 2011-02, Vol.30 (2), p.201-208
Main Authors: Szabados, F, Woloszyn, J, Kaase, M, Gatermann, S. G
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The relative sensitivity of commercial agglutination kits for fast identification of S. aureus is usually given to be about 98%. This reported sensitivity has sometimes been questioned. In this study, three collections of molecularly defined, single-copy strains of S. aureus were used to compare the sensitivities of agglutination-based identification and the MALDI-TOF mass spectrometry-based identification using the Biotyper 2.0 database to a molecularly defined reference method. Clinical isolates (n = 363) of methicillin-susceptible S. aureus (MSSA) and 240 clinical isolates of methicillin-resistant S. aureus (MRSA) were included. In order to rule out a predominance of local MRSA-strains, a collection of 104 pulsed-field-gel electrophoresis divergent MRSA strains were also tested. MALDI-TOF MS using Biotyper database (Bruker) identified all isolates, whereas the Slidex Staph Plus (bioMérieux) detected only 98.0% of the MSSA, 94.5% of the MRSA and only 70.1% of the MRSA of the molecularly divergent strain collection. Interestingly, strains with a false-negative test result in the agglutination methods were mostly spa-type t001 and t001 related. The MALDI-TOF MS based identification can thus be used as an alternative identification method for suspected false-negative results from the agglutination tests, especially if the local prevalence of t001 and t001 related strains is high.
ISSN:0934-9723
1435-4373
DOI:10.1007/s10096-010-1070-4