Sweat analysis using indirect ion-selective electrode on the routine chemistry analyser meets UK guidelines
Background According to consensus documents, sweat testing remains the standard for diagnosing cystic fibrosis. We studied a novel method of sweat analysis on two chemistry platforms: Dimension Vista (Siemens Healthcare Diagnostics) and Aeroset (Abbott). Methods Sweat collected from healthy individu...
Saved in:
Published in: | Annals of clinical biochemistry 2011-07, Vol.48 (4), p.374-376 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
According to consensus documents, sweat testing remains the standard for diagnosing cystic fibrosis. We studied a novel method of sweat analysis on two chemistry platforms: Dimension Vista (Siemens Healthcare Diagnostics) and Aeroset (Abbott).
Methods
Sweat collected from healthy individuals and aqueous samples in the critical range for sweat analysis were measured using the Orion 290 plus (Thermo Electron Corporation), Chiron 925 chloride analyser (Sherwood Scientific Ltd), flame photometry, sweat-chek analyser and indirect ion-selective electrodes (ISE) of Vista and Aeroset. For measurement on ISE, an addition protocol was designed.
Results
Within-run and between-run variation coefficients of Aeroset and Vista remained |
---|---|