Beryllium in urine by ICP-MS: a comparison of low level exposed workers and unexposed persons
Purpose To develop a sensitive and reproducible method for urinary beryllium and to use this method to establish levels in workers at an aluminium smelter and in unexposed persons. Methods A method was developed for urinary beryllium using a Thermo ICP-MS Series 1, which was used to determine beryll...
Saved in:
Published in: | International archives of occupational and environmental health 2011-08, Vol.84 (6), p.697-704 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose
To develop a sensitive and reproducible method for urinary beryllium and to use this method to establish levels in workers at an aluminium smelter and in unexposed persons.
Methods
A method was developed for urinary beryllium using a Thermo ICP-MS Series 1, which was used to determine beryllium concentrations in urine from 62 people with no known occupational exposure to beryllium and 167 workers with potential exposure to beryllium at an aluminium smelter, where beryllium exists as an impurity in the bauxite ore.
Results
The analytical method has a detection limit (based on three times the background equivalent concentration of the blank) for beryllium in urine of 6 ng/L. The mean and 90th percentiles of urinary beryllium for workers were 19.5 and 42.0 ng/L and compared with 11.6 and 20.0 ng/L in people not occupationally exposed to beryllium. Statistical analysis using mixed effects models showed that workers had 47% (in 135 paired samples) higher levels of urinary beryllium at the end of the working week compared to the start of week and that the workers who smoked also had significantly higher levels of urinary beryllium compared to those that did not smoke. There was also a statistically significant difference between workers and controls in urinary beryllium concentrations not corrected for creatinine.
Conclusions
A sensitive and reliable analytical method was developed for urinary beryllium by ICP-MS. The workers in this study were exposed to beryllium at very low levels. |
---|---|