Loading…

Effects of Environmental Conditions on the Performance of Thermal Imagers

An experimental setup for characterization of thermal imagers has been designed and constructed. The setup comprises a climatic chamber and a temperature-controlled flat blackbody target that allow testing of thermal imagers in the ambient temperature range from −13 °C to + 23  °C and at target temp...

Full description

Saved in:
Bibliographic Details
Published in:International journal of thermophysics 2011, Vol.32 (1-2), p.248-257
Main Authors: Vendt, R., Juurma, M., Jaanson, P., Vabson, V., Kübarsepp, T., Noorma, M.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An experimental setup for characterization of thermal imagers has been designed and constructed. The setup comprises a climatic chamber and a temperature-controlled flat blackbody target that allow testing of thermal imagers in the ambient temperature range from −13 °C to + 23  °C and at target temperatures from −15 °C to + 120 °C. The setup has been tested with a high quality thermal imager for scientific use and a mid-range thermal imager for industrial applications. The test results indicate that the setup allows monitoring the changes in indication of thermal imagers within the standard uncertainty limits of 0.5 °C.
ISSN:0195-928X
1572-9567