Loading…
Effects of Environmental Conditions on the Performance of Thermal Imagers
An experimental setup for characterization of thermal imagers has been designed and constructed. The setup comprises a climatic chamber and a temperature-controlled flat blackbody target that allow testing of thermal imagers in the ambient temperature range from −13 °C to + 23 °C and at target temp...
Saved in:
Published in: | International journal of thermophysics 2011, Vol.32 (1-2), p.248-257 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An experimental setup for characterization of thermal imagers has been designed and constructed. The setup comprises a climatic chamber and a temperature-controlled flat blackbody target that allow testing of thermal imagers in the ambient temperature range from −13 °C to + 23 °C and at target temperatures from −15 °C to + 120 °C. The setup has been tested with a high quality thermal imager for scientific use and a mid-range thermal imager for industrial applications. The test results indicate that the setup allows monitoring the changes in indication of thermal imagers within the standard uncertainty limits of 0.5 °C. |
---|---|
ISSN: | 0195-928X 1572-9567 |
DOI: | 10.1007/s10765-010-0896-4 |