Loading…

NEMA count-rate evaluation of the first and second generation of the Ecat Exact and Ecat Exact HR family of scanners

The first and second generation of the Exact and Exact HR family of scanners has been evaluated in terms of noise equivalent count rate (NEC) and count-rate capabilities. The new National Electrical Manufacturers Association standard was used for the evaluation. In spite of improved electronics and...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on nuclear science 2002-06, Vol.49 (3), p.640-643
Main Authors: Eriksson, L., Wienhard, K., Eriksson, M., Casey, M.E., Knoess, C., Bruckbauer, T., Hamill, J., Mulnix, T., Vollmar, S., Bendriem, B., Heiss, W.D., Nutt, R.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The first and second generation of the Exact and Exact HR family of scanners has been evaluated in terms of noise equivalent count rate (NEC) and count-rate capabilities. The new National Electrical Manufacturers Association standard was used for the evaluation. In spite of improved electronics and improved count-rate capabilities, the peak NEC was found to be fairly constant between the generations. The results are discussed in terms of the different electronic solutions for the two generations and its implications on system dead time and NEC count-rate capability.
ISSN:0018-9499
1558-1578
DOI:10.1109/TNS.2002.1039542