Failure of HA coating on a gritblasted acetabular cup: 155 patients followed for 7-10 years
We report the outcome of 191 acetabular gritblasted titanium cups with a hemispherical design for press-fit insertion and coated with hydroxyapatite. The prosthesis was made of gritblasted titanium entirely coated with hydroxyapatite. 155 patients aged 15-78 years were operated on during the years 1...
Saved in:
Published in: | Acta orthopaedica 2002, Vol.73 (1), p.104-108 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report the outcome of 191 acetabular gritblasted titanium cups with a hemispherical design for press-fit insertion and coated with hydroxyapatite. The prosthesis was made of gritblasted titanium entirely coated with hydroxyapatite. 155 patients aged 15-78 years were operated on during the years 1991-1993 and followed for 7-10 years. During this period, 39 cups were revised because of mechanical loosening, a further 9 had radiolucent lines and 2 focal osteolysis. None of these 11 patients had clinical symptoms. Failure was associated with age, wear and radiolucency/osteolysis. At revision, we found that the soft tissues were discolored, and that most of the coating had disappeared. This design of hydroxyapatite-coated cups has a high rate of debonding and failure. |
---|---|
ISSN: | 1745-3674 0001-6470 1745-3682 |
DOI: | 10.1080/000164702317281503 |