Image distortion in MRI-based polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems
We have studied effects of MR (magnetic resonance) image distortion on polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems. MR images of BANG polymer gel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by...
Saved in:
Published in: | Medical physics (Lancaster) 2002-05, Vol.29 (5), p.797-802 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have studied effects of MR (magnetic resonance) image distortion on polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems. MR images of BANG polymer gel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by varying the directions of frequency encoding and the receiver bandwidth. The phantoms were also CT (computed tomography) scanned. The studies showed that the measured dose distributions are shifted by
1.8±0.5
mm
(2 pixels) in the frequency encoding direction. The magnitude of the shift is inversely proportional to the receiver bandwidth in agreement with theory. Comparison of MRI with CT showed the same image shift. We concluded that the discrepancy is caused by MR image distortion due to a difference in susceptibility effects between the phantom and the fiducial markers of the Leksell localization box. |
---|---|