A TL/OSL Emission Spectrometer Extension of the Risø Reader
The integration of a low cost, compact sized spectrometer with the Risø reader is described. The luminescence light emitted by the sample is transmitted by an optical fibre onto a fixed entrance slit of a spectrograph. The light is measured with a high sensitivity 2048-element CCD-linear array detec...
Saved in:
Published in: | Radiation protection dosimetry 2002-01, Vol.101 (1-4), p.111-114 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The integration of a low cost, compact sized spectrometer with the Risø reader is described. The luminescence light emitted by the sample is transmitted by an optical fibre onto a fixed entrance slit of a spectrograph. The light is measured with a high sensitivity 2048-element CCD-linear array detector (Avantes PC2000) sensitive in the UV-VIS region. The CCD array has a low readout noise and a photon sensitivity of 86 photons per count. Examples are given of both thermally and optically stimulated 3-D spectra showing the luminescence spectra as a function of temperature and time respectively. Spectra from CaF2:Tm (TLD-300) chips could be distinguished from the background (2 SD) at a 0.7 Gy dose level. |
---|---|