Experimental validation of a commercial 3D dose verification system for intensity-modulated arc therapies

We validate the dosimetric performance of COMPASS®, a novel 3D quality assurance system for verification of volumetric-modulated arc therapy (VMAT) treatment plans that can correlate the delivered dose to the patient's anatomy, taking into account the tissue inhomogeneity. The accuracy of treat...

Full description

Saved in:
Bibliographic Details
Published in:Physics in medicine & biology 2010-10, Vol.55 (19), p.5619-5633
Main Authors: Boggula, Ramesh, Lorenz, Friedlieb, Mueller, Lutz, Birkner, Mattias, Wertz, Hansjoerg, Stieler, Florian, Steil, Volker, Lohr, Frank, Wenz, Frederik
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We validate the dosimetric performance of COMPASS®, a novel 3D quality assurance system for verification of volumetric-modulated arc therapy (VMAT) treatment plans that can correlate the delivered dose to the patient's anatomy, taking into account the tissue inhomogeneity. The accuracy of treatment delivery was assessed by the COMPASS® for 12 VMAT plans, and the resulting assessments were evaluated using an ionization chamber and film measurements. Dose-volume relationships were evaluated by the COMPASS® for three additional treatment plans and these were used to verify the accuracy of treatment planning dose calculations. The results matched well between COMPASS® and measurements for the ionization chamber (≤3%) and film (73-99% for gamma((3%/3 mm)) < 1 and 98-100% for gamma((5%/5 mm)) < 1) for the phantom plans. Differences in dose-volume statistics for the average dose to the PTV were within 2.5% for three treatment plans. For the structures located in the low-dose region, a maximum difference of
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/55/19/001