Measurement of backscatter factors for low energy radiotherapy (0.1-2.0 mm Al HVL) using thermoluminescence dosimetry

Significant discrepancies of up to 10% exist between backscatter factors (BSF) recommended in a recent IAEA dosimetry Code of Practice (1987) compared with those published in Br. J. Radiol. Supplement 17 (1983), for the x-ray quality range 0.1-2.0 mm Al HVL. In an attempt to resolve this discrepancy...

Full description

Saved in:
Bibliographic Details
Published in:Physics in medicine & biology 1990-09, Vol.35 (9), p.1247-1254
Main Authors: Harrison, R M, Walker, C, Aukett, R J
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Significant discrepancies of up to 10% exist between backscatter factors (BSF) recommended in a recent IAEA dosimetry Code of Practice (1987) compared with those published in Br. J. Radiol. Supplement 17 (1983), for the x-ray quality range 0.1-2.0 mm Al HVL. In an attempt to resolve this discrepancy, BSFs have been measured using thermoluminescence dosimetry (TLD) with small lithium borate chips, in order to minimise displacement effects associated with the use of larger volume ionisation chambers. Although subject to uncertainties inherent in the TLD calibration and readout process, the results indicate that the BJR (1983) data overestimate BSFs in this quality range. Broad agreement with the IAEA data is indicated, for the limited number of x-ray qualities and field sizes used.
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/35/9/005