A 180-V pp Integrated Linear Amplifier for Ultrasonic Imaging Applications in a High-Voltage CMOS SOI Technology

This brief presents a monolithically integrated fully differential linear HV amplifier as the driver of an ultrasonic transducer. The linear amplifier is capable of transmitting HV arbitrary signals with a very low harmonic distortion, which is suitable for tissue harmonic imaging and other ultrason...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on circuits and systems. II, Express briefs Express briefs, 2015-02, Vol.62 (2), p.149
Main Authors: Sun, Kexu, Gao, Zheng, Gui, Ping, Wang, Rui, Oguzman, Ismail, Xu, Xiaochen, Vasanth, Karthik, Zhou, Qifa, Shung, K Kirk
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This brief presents a monolithically integrated fully differential linear HV amplifier as the driver of an ultrasonic transducer. The linear amplifier is capable of transmitting HV arbitrary signals with a very low harmonic distortion, which is suitable for tissue harmonic imaging and other ultrasonic modes for enhanced imaging quality. The amplifier is designed and implemented using the 0.7- m CMOS silicon-on-insulator process with 120-V devices. The amplifier, when driving a load of 300 pF in parallel with 100 Ω, is capable of transmitting a sine-wave signal with a frequency of up to 4.4 MHz, a maximum signal swing of 180 V , and a second-order harmonic distortion (HD2) of -56 dBc but only dissipating an average power of 62 mW with a 0.1% duty cycle.