Loading…

Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring

The present study investigated the relationship between blood and subcutaneous interstitial fluid (ISF) glucose by employing an amperometric glucose sensor specifically developed for 3-day continuous glucose monitoring. The apparent sensor sensitivity and ISF glucose equilibration delay were estimat...

Full description

Saved in:
Bibliographic Details
Published in:American journal of physiology: endocrinology and metabolism 1999-09, Vol.277 (3), p.E561
Main Authors: Rebrin, Kerstin, Steil, Garry M, van Antwerp, William P, Mastrototaro, John J
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The present study investigated the relationship between blood and subcutaneous interstitial fluid (ISF) glucose by employing an amperometric glucose sensor specifically developed for 3-day continuous glucose monitoring. The apparent sensor sensitivity and ISF glucose equilibration delay were estimated on separate days during hyperglycemic clamps in four dogs in which insulin was either suppressed with somatostatin, allowed to change, or increased with an exogenous infusion. A 2-h sensor "settling-in" period was allowed before the clamps. During insulin deficiency, the sensor sensitivity and ISF glucose delay were 0.23 ± 0.03 nA per mg/dl and 4.4 ± 0.8 min. Sensitivity was not affected by increases in endogenous (0.30 ± 0.04 vs. 0.28 ± 0.04 nA per mg/dl) or exogenous insulin (0.18 ± 0.01 vs. 0.16 ± 0.01 nA per mg/dl) nor was the delay (3.3 ± 1.2 vs. 5.7 ± 1.1 and 9.2 ± 2.6 vs. 12.3 ± 1.7 min; P > 0.05 for all). Sensor glucose accurately predicted plasma glucose without correcting for delays 0.9 for all), whereas for longer delays a digital corrective filter was used ( r = 0.91 with filter). We conclude that the relationship between blood and ISF glucose is not affected by insulin and that delays in ISF glucose equilibration can be corrected with digital filters.
ISSN:1522-1555