Loading…

Comprehensive Evaluation of B 1 + -corrected FISP-based Magnetic Resonance Fingerprinting: Accuracy, Repeatability and Reproducibility of T 1 and T 2 Relaxation Times for ISMRM/NIST System Phantom and Volunteers

This study aimed to evaluate comprehensively; accuracy, repeatability and reproducibility of T and T relaxation times measured by magnetic resonance fingerprinting using B -corrected fast imaging with steady-state precession (FISP-MRF). The International Society of Magnetic Resonance in Medicine/Nat...

Full description

Saved in:
Bibliographic Details
Published in:Magnetic resonance in medical sciences 2020-01
Main Authors: Kato, Yutaka, Ichikawa, Kazushige, Okudaira, Kuniyasu, Taoka, Toshiaki, Kawaguchi, Hirokazu, Murata, Katsutoshi, Maruyama, Katsuya, Koerzdoerfer, Gregor, Pfeuffer, Josef, Nittka, Mathias, Naganawa, Shinji
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This study aimed to evaluate comprehensively; accuracy, repeatability and reproducibility of T and T relaxation times measured by magnetic resonance fingerprinting using B -corrected fast imaging with steady-state precession (FISP-MRF). The International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) phantom was scanned for 100 days, and six healthy volunteers for 5 days using a FISP-MRF prototype sequence. Accuracy was evaluated on the phantom by comparing relaxation times measured by FISP-MRF with the reference values provided by the phantom manufacturer. Daily repeatability was characterized as the coefficient of variation (CV) of the measurements over 100 days for the phantom and over 5 days for volunteers. In addition, the cross-scanner reproducibility was evaluated in volunteers. In the phantom study, T and T values from FISP-MRF showed a strong linear correlation with the reference values of the phantom (R = 0.9963 for T ; R = 0.9966 for T ). CVs were
ISSN:1880-2206
DOI:10.2463/mrms.mp.2019-0016