Monolithic 3D micromixer with an impeller for glass microfluidic systems
The performance of micromixers, namely their mixing efficiency and throughput, is a critical component in increasing the overall efficiency of microfluidic systems ( e.g. , lab-on-a-chip and μ-TAS). Most previously reported high-performance micromixers use active elements with some external power to...
Saved in:
Published in: | Lab on a chip 2020-11, Vol.2 (23), p.4474-4485 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The performance of micromixers, namely their mixing efficiency and throughput, is a critical component in increasing the overall efficiency of microfluidic systems (
e.g.
, lab-on-a-chip and μ-TAS). Most previously reported high-performance micromixers use active elements with some external power to induce turbulence, or contain long and complex fluidic channels with obstacles to increase diffusion. In this paper, we introduce a new type of 3D impeller micromixer built within a single fused silica substrate. The proposed device is composed of microchannels with three inlets and a tank, with a mixing impeller passively rotated by axial flow. The passive micromixer is directly fabricated inside a glass plate using a selective laser-induced etching technique. The mixing tank, with its rotating shaft and 3D pitched blade impeller, exists within a micro-cavity with a volume of only 0.28 mm
3
. A mixing efficiency of 99% is achieved in mixing experiments involving three dye colours over flow rates ranging from 1.5-30 mL min
−1
, with the same flow rates also applied to a sodium hydroxide-based bromothymol blue indicator and a hydrochloric acid chemical solution. To verify the reliable performance of the proposed device, we compare the mixing index with a general self-circulation-type chamber mixer to demonstrate the improved mixing efficiency achieved by rotating the impeller. No cracking or breakage of the device is observed under high inner pressures or when the maximum flow rate is applied to the mixer. The proposed microfluidic system based on a compact built-in 3D micromixer with an impeller opens the door to robust, highly efficient, and high-throughput glass-based platforms for micro-centrifuges, cell sorters, micro-turbines, and micro-pumps.
We introduce a new 3D impeller micromixer built within a single glass substrate using ultrafast laser process and it shows high mixing efficiency up to 99% and throughput of 30 mL min
−1
with a short mixing channel length of 0.98 mm. |
---|---|
ISSN: | 1473-0197 1473-0189 |
DOI: | 10.1039/d0lc00823k |