Loading…

Cold neutron radiation dose effects on a 6 LiF:ZnS(Ag) neutron detector with wavelength shifting fibers and SiPM photodetector

A LiF:ZnS(Ag) based cold neutron detector with wavelength shifting (WLS) fibers and SiPM photodetector was developed at the NIST Center for Neutron Research for the CANDoR instrument (Chromatic Analysis Neutron Diffractometer or Reflectometer). A series of detectors were irradiated with neutron dose...

Full description

Saved in:
Bibliographic Details
Published in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2021-01, Vol.986
Main Authors: Pritchard, K, Hussey, D, Osovizky, A, Ziegler, J, Binkley, E, Tsai, P, Hadad, N, Jackson, M, Hurlbut, C, Baltic, G M, Majkrzak, C F, Maliszewskyj, N C
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A LiF:ZnS(Ag) based cold neutron detector with wavelength shifting (WLS) fibers and SiPM photodetector was developed at the NIST Center for Neutron Research for the CANDoR instrument (Chromatic Analysis Neutron Diffractometer or Reflectometer). A series of detectors were irradiated with neutron doses ranging between 1E+11 n/cm to 6E+12 n/cm . It was found that the neutron absorbing Li isotope was not measurably depleted, but the photonic yield of the detector deteriorated with increasing neutron dose. Photonic yields were compared before and after neutron exposure by comparing pulse energy spectrum photopeaks before and after exposure. A typical detector used in the CANDoR instrument is expected to withstand a cumulative cold neutron dose of at least 1E+12 n/cm before degrading to an unfit state. The component parts of the detector could not be separated, so the degradation of the bulk scintillator and WLS fibers could not be gauged separately.
ISSN:0168-9002