Radiation dose considerations in digital radiography with an anti‐scatter grid: A study using adult and pediatric phantoms

Background When using an anti‐scatter grid, a decrease in receptor dose caused by its X‐ray absorption seems to lead to the misperception that radiation dose needs to be increased even in digital radiography (DR). Objective To demonstrate that there is no need to increase radiation dose in DR with a...

Full description

Saved in:
Bibliographic Details
Published in:Journal of applied clinical medical physics 2023-09, Vol.24 (9), p.e14081-n/a
Main Authors: Kawashima, Hiroki, Ichikawa, Katsuhiro, Kitao, Azusa, Matsubara, Takashi, Sugiura, Takumi, Kobayashi, Tomohiro, Kobayashi, Satoshi
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background When using an anti‐scatter grid, a decrease in receptor dose caused by its X‐ray absorption seems to lead to the misperception that radiation dose needs to be increased even in digital radiography (DR). Objective To demonstrate that there is no need to increase radiation dose in DR with a grid, based on a visual evaluation using an adult and a pediatric abdomen phantom (PAD and PPD, respectively). Materials and methods Phantom images with and without a grid were obtained with exposure parameters determined based on a preliminarily measured signal‐to‐noise ratio improvement factor (SIF), an index for potential dose reduction when using a grid. In visual evaluation, four radiologists compared phantom images with a grid applied at different dose reduction rates (0% [no reduction], 18%, 36%, and 59% for PAD and 0% and 11% for PPD) against an image without a grid at the baseline dose (as the reference). They graded the overall image quality of the former relative to that of the latter (reference) on a 3‐point scale (3 = better, 2 = almost equal, 1 = worse). Results The mean scores for dose reduction rates of 0%, 18%, 36%, and 59% were 3.00, 3.00, 2.75, and 1.00, respectively, for PAD; those for 0% and 11% were 2.13 and 1.63, respectively, for PPD. These results support the validity of our view that no dose increase is necessary when using an anti‐scatter grid. Actually, there is even a potential for improvement in image quality with dose reduction rates of ≤36% for PAD. Conclusion It is worth reconsidering the necessity of increasing radiation dose in the DR imaging of the adult and pediatric abdomens with an anti‐scatter grid.
ISSN:1526-9914
1526-9914
DOI:10.1002/acm2.14081