Design and Development of a Clot Burst Pressure Device to Investigate Resuscitation Strategies
A reduction in clot strength is a hallmark feature of trauma-induced coagulopathy. A better understanding of clot integrity can optimize resuscitation strategies. We designed a device to gauge clot strength by pressurizing fluids over a formed clot and measuring the pressure needed to dislodge the c...
Saved in:
Published in: | The Journal of surgical research 2023-11, Vol.291, p.646-652 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A reduction in clot strength is a hallmark feature of trauma-induced coagulopathy. A better understanding of clot integrity can optimize resuscitation strategies. We designed a device to gauge clot strength by pressurizing fluids over a formed clot and measuring the pressure needed to dislodge the clot. We hypothesized that this device could distinguish between clots formed in hypocoagulable and hypercoagulable states by observing differences in the clot burst pressure.
Whole blood from healthy volunteers was collected into sodium citrate tubes and was treated with heparin or fibrinogen to generate clots in a hypocoagulable or hypercoagulable state, respectively. Small bore holes were drilled into polystyrene plates, and recalcified blood was pipetted into the holes. Plates were incubated at 37°C for 30 min to form clots. A pressure cap with an inlet for fluid from a syringe pump and an outlet leading to a measurement column was secured in the wells with a watertight seal.
Clot burst pressure was normalized to individual baseline values to account for inherent differences in clot strength. The 1.0 g/L and 2.0 g/L fibrinogen groups were 1.65 ± 0.07 (P = 0.0078) and 2.26 ± 0.16 (P = 0.0078) times as strong as baseline, respectively. The 0.10, 0.15, or 0.20 USP units/mL groups were 0.388 ± 0.07 (P = 0.125), 0.31 ± 0.07 (P = 0.125), 0.21 ± 0.07 (P = 0.125) times as strong as baseline, respectively. Data were analyzed using Wilcoxon matched pairs signed rank testing.
This device tests clot strength using burst pressure, an easily interpreted clinical parameter not measured in existing devices. Future work can test blood from trauma patients to better understand trauma pathophysiology. |
---|---|