Loading…
Real-time tumor-tracking radiotherapy with SyncTraX for primary liver tumors requiring isocenter shift
Abstract The SyncTraX series enables real-time tumor-tracking radiotherapy through the real-time recognition of a fiducial marker using fluoroscopic images. In this system, the isocenter should be located within approximately 5–7.5 cm from the marker, depending on the version, owing to the limited f...
Saved in:
Published in: | Journal of radiation research 2024-01, Vol.65 (1), p.92-99 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract
The SyncTraX series enables real-time tumor-tracking radiotherapy through the real-time recognition of a fiducial marker using fluoroscopic images. In this system, the isocenter should be located within approximately 5–7.5 cm from the marker, depending on the version, owing to the limited field of view. If the marker is placed away from the tumor, the isocenter should be shifted toward the marker. This study aimed to investigate stereotactic body radiotherapy (SBRT) outcomes of primary liver tumors treated with SyncTraX in cases where the isocenter was shifted marginally or outside the planning target volume (PTV). Twelve patients with 13 liver tumors were included in the analysis. Their isocenter was shifted toward the marker and was placed marginally or outside the PTV. The prescribed doses were generally 40 Gy in four fractions or 48 Gy in eight fractions. The overall survival (OS) and local control (LC) rates were calculated using the Kaplan–Meier method. All patients completed the scheduled SBRT. The median distance between the fiducial marker and PTV centroid was 56.0 (interquartile range [IQR]: 52.7–66.7) mm. By shifting the isocenter toward the marker, the median distance between the marker and isocenter decreased to 34.0 (IQR: 33.4–39.7) mm. With a median follow-up period of 25.3 (range: 6.9–70.0) months, the 2-year OS and LC rates were 100.0% (95% confidence interval: 100–100). An isocenter shift makes SBRT with SyncTraX feasible in cases where the fiducial marker is distant from the tumor. |
---|---|
ISSN: | 0449-3060 1349-9157 1349-9157 |
DOI: | 10.1093/jrr/rrad088 |