A standardised communication tool reduces radiation exposure associated with intraoperative fluoroscopy
Background The widespread use of intraoperative fluoroscopy in orthopaedic procedures has revolutionised surgical practice. However, there are risks associated with using ionising radiation. Efforts to reduce radiation exposure include low-dose imaging protocols and lead protective equipment. Curren...
Saved in:
Published in: | Irish journal of medical science 2024-02, Vol.193 (1), p.257-263 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
The widespread use of intraoperative fluoroscopy in orthopaedic procedures has revolutionised surgical practice. However, there are risks associated with using ionising radiation. Efforts to reduce radiation exposure include low-dose imaging protocols and lead protective equipment. Current communication during fluoroscopic procedures can be inefficient and lead to excessive radiation exposure for patients and staff.
Aims
This study aims to implement a communication tool with standardised commands to reduce radiation exposure in an Irish orthopaedic department.
Methods
Radiation exposure was evaluated using dose-area product (DAP) measured in uGy/m2. A control group was recorded before implementing the communication tool. Training sessions were conducted and posters of the standardised commands were displayed. Feedback was collected from surgeons and radiographers via surveys. Statistical analysis was performed to compare pre- and post-intervention groups.
Results
A total of 673 surgical cases were included over 6 months. The post-intervention group showed a mean reduction in radiation exposure from 59.8 to 36.4 uGy/m2 (
p
|
---|---|
ISSN: | 0021-1265 1863-4362 1863-4362 |
DOI: | 10.1007/s11845-023-03442-2 |