Loading…

Development of a Slow-Degrading Polymerized Curcumin Coating for Intracortical Microelectrodes

Intracortical microelectrodes are used with brain–computer interfaces to restore lost limb function following nervous system injury. While promising, recording ability of intracortical microelectrodes diminishes over time due, in part, to neuroinflammation. As curcumin has demonstrated neuroprotecti...

Full description

Saved in:
Bibliographic Details
Published in:ACS applied bio materials 2023-02, Vol.6 (2), p.806-818
Main Authors: Ziemba, Alexis M., Woodson, Mary Clare Crochiere, Funnell, Jessica L., Wich, Douglas, Balouch, Bailey, Rende, Deniz, Amato, Dahlia N., Bao, Jonathan, Oprea, Ingrid, Cao, Dominica, Bajalo, Neda, Ereifej, Evon S., Capadona, Jeffrey R., Palermo, Edmund F., Gilbert, Ryan J.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Intracortical microelectrodes are used with brain–computer interfaces to restore lost limb function following nervous system injury. While promising, recording ability of intracortical microelectrodes diminishes over time due, in part, to neuroinflammation. As curcumin has demonstrated neuroprotection through anti-inflammatory activity, we fabricated a 300 nm-thick intracortical microelectrode coating consisting of a polyurethane copolymer of curcumin and polyethylene glycol (PEG), denoted as poly­(curcumin-PEG1000 carbamate) (PCPC). The uniform PCPC coating reduced silicon wafer hardness by two orders of magnitude and readily absorbed water within minutes, demonstrating that the coating is soft and hydrophilic in nature. Using an in vitro release model, curcumin eluted from the PCPC coating into the supernatant over 1 week; the majority of the coating was intact after an 8-week incubation in buffer, demonstrating potential for longer term curcumin release and softness. Assessing the efficacy of PCPC within a rat intracortical microelectrode model in vivo, there were no significant differences in tissue inflammation, scarring, neuron viability, and myelin damage between the uncoated and PCPC-coated probes. As the first study to implant nonfunctional probes with a polymerized curcumin coating, we have demonstrated the biocompatibility of a PCPC coating and presented a starting point in the design of poly­(pro-curcumin) polymers as coating materials for intracortical electrodes.
ISSN:2576-6422
2576-6422
DOI:10.1021/acsabm.2c00969