Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts
This study was designed to determine the correlation between heamatological parameters by Sysmex KX-21N automated hematology analyzer with the manual methods. Sixty (60) subjects were randomly selected from both apparently healthy subjects and those who have different blood disorders from the Univer...
Saved in:
Published in: | BMC clinical pathology 2010-04, Vol.10 (1), p.3-3, Article 3 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113 |
---|---|
cites | cdi_FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113 |
container_end_page | 3 |
container_issue | 1 |
container_start_page | 3 |
container_title | BMC clinical pathology |
container_volume | 10 |
creator | Ike, Samuel O Nubila, Thomas Ukaejiofo, Ernest O Nubila, Imelda N Shu, Elvis N Ezema, Ifeyinwa |
description | This study was designed to determine the correlation between heamatological parameters by Sysmex KX-21N automated hematology analyzer with the manual methods.
Sixty (60) subjects were randomly selected from both apparently healthy subjects and those who have different blood disorders from the University of Teaching Hospital (UNTH), Ituku-Ozalla, Enugu, Enugu State, Nigeria. Three (3)mls of venous blood sample was collected aseptically from each subject into tri-potassium ethylenediamine tetra-acetic acid (K3EDTA) for the analysis of haematological parameters using the automated and the manual methods.
The blood film report by the manual method showed that 50% of the subjects were normocytic-normochromic while the other 50% revealed different abnormal blood pictures. Also, there were statistically significant differences (p < 0.05) in mean cell hemoglobin concentrations (MCHC) between the two methods. Similarly, the mean (S.E) values of hemoglobin, packed cell volume, platelet and total white cell counts demonstrated statistically significant difference (p < 0.001) and correlated positively when both methods were compared.
From the present study, it can be concluded that the automated hematology analyzer readings correlated well with readings by the standard manual method, although the latter method gave additional diagnostic information on the blood pictures. While patients' care and laboratory operations could be optimized by using manual microscopic examination as a reflective substitute for automated methods, usage of automated method would ease our workload and save time for patients. |
doi_str_mv | 10.1186/1472-6890-10-3 |
format | article |
fullrecord | <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_2873444</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A226839134</galeid><sourcerecordid>A226839134</sourcerecordid><originalsourceid>FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113</originalsourceid><addsrcrecordid>eNp1kk2LFDEQhoMo7rp69SgBD556zdd00hdhGPxYXPSggreQTlfPRDrJmHSL7cHfbtpZx11cyaGKqrce3hSF0GNKzilV9XMqJKtq1ZCKkorfQafHwt1r-Ql6kPMXQqhUlN1HJ4wIWpNanaKfm-j3JrkcA4493hnwZoxD3DprBlw6xsMIKeNuCd4F6HA743EH-MOcPXzHbz_jCrOLd9hMYyzDRfCXMmMTzDD_gFSS7veYN2EqaBunMOaH6F5vhgyPruIZ-vTq5cfNm-ry_euLzfqyaldUjJVsGy6VJEoxYKQRlIve1g0RDNqmsRY6Rluwsparpu0oh46WbwtB5IqImlJ-hl4cuPup9dBZCGMyg94n502adTRO3-wEt9Pb-E0zJbkQogDWB0Dr4n8ANzs2er3sXy_715RoXhjPrkyk-HWCPGrvsoVhMAHilLXkfCWkkKoonx6UWzOAdqGPhWkXtV4zVivelA0U1fktqvI68M7GAL0r9dsGbIo5J-iP_ou_5Zz-dfzk-tqO8j_3w38B8TnGBA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>733547478</pqid></control><display><type>article</type><title>Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts</title><source>ProQuest - Publicly Available Content Database</source><source>PubMed Central</source><creator>Ike, Samuel O ; Nubila, Thomas ; Ukaejiofo, Ernest O ; Nubila, Imelda N ; Shu, Elvis N ; Ezema, Ifeyinwa</creator><creatorcontrib>Ike, Samuel O ; Nubila, Thomas ; Ukaejiofo, Ernest O ; Nubila, Imelda N ; Shu, Elvis N ; Ezema, Ifeyinwa</creatorcontrib><description>This study was designed to determine the correlation between heamatological parameters by Sysmex KX-21N automated hematology analyzer with the manual methods.
Sixty (60) subjects were randomly selected from both apparently healthy subjects and those who have different blood disorders from the University of Teaching Hospital (UNTH), Ituku-Ozalla, Enugu, Enugu State, Nigeria. Three (3)mls of venous blood sample was collected aseptically from each subject into tri-potassium ethylenediamine tetra-acetic acid (K3EDTA) for the analysis of haematological parameters using the automated and the manual methods.
The blood film report by the manual method showed that 50% of the subjects were normocytic-normochromic while the other 50% revealed different abnormal blood pictures. Also, there were statistically significant differences (p < 0.05) in mean cell hemoglobin concentrations (MCHC) between the two methods. Similarly, the mean (S.E) values of hemoglobin, packed cell volume, platelet and total white cell counts demonstrated statistically significant difference (p < 0.001) and correlated positively when both methods were compared.
From the present study, it can be concluded that the automated hematology analyzer readings correlated well with readings by the standard manual method, although the latter method gave additional diagnostic information on the blood pictures. While patients' care and laboratory operations could be optimized by using manual microscopic examination as a reflective substitute for automated methods, usage of automated method would ease our workload and save time for patients.</description><identifier>ISSN: 1472-6890</identifier><identifier>EISSN: 1472-6890</identifier><identifier>DOI: 10.1186/1472-6890-10-3</identifier><identifier>PMID: 20416068</identifier><language>eng</language><publisher>England: BioMed Central Ltd</publisher><subject>Blood ; Blood diseases ; Demographic aspects ; Diagnosis ; Medical examination ; Methods ; Parameter estimation ; Research article</subject><ispartof>BMC clinical pathology, 2010-04, Vol.10 (1), p.3-3, Article 3</ispartof><rights>COPYRIGHT 2010 BioMed Central Ltd.</rights><rights>Copyright ©2010 Ike et al; licensee BioMed Central Ltd. 2010 Ike et al; licensee BioMed Central Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113</citedby><cites>FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2873444/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2873444/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,27923,27924,37012,53790,53792</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/20416068$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ike, Samuel O</creatorcontrib><creatorcontrib>Nubila, Thomas</creatorcontrib><creatorcontrib>Ukaejiofo, Ernest O</creatorcontrib><creatorcontrib>Nubila, Imelda N</creatorcontrib><creatorcontrib>Shu, Elvis N</creatorcontrib><creatorcontrib>Ezema, Ifeyinwa</creatorcontrib><title>Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts</title><title>BMC clinical pathology</title><addtitle>BMC Clin Pathol</addtitle><description>This study was designed to determine the correlation between heamatological parameters by Sysmex KX-21N automated hematology analyzer with the manual methods.
Sixty (60) subjects were randomly selected from both apparently healthy subjects and those who have different blood disorders from the University of Teaching Hospital (UNTH), Ituku-Ozalla, Enugu, Enugu State, Nigeria. Three (3)mls of venous blood sample was collected aseptically from each subject into tri-potassium ethylenediamine tetra-acetic acid (K3EDTA) for the analysis of haematological parameters using the automated and the manual methods.
The blood film report by the manual method showed that 50% of the subjects were normocytic-normochromic while the other 50% revealed different abnormal blood pictures. Also, there were statistically significant differences (p < 0.05) in mean cell hemoglobin concentrations (MCHC) between the two methods. Similarly, the mean (S.E) values of hemoglobin, packed cell volume, platelet and total white cell counts demonstrated statistically significant difference (p < 0.001) and correlated positively when both methods were compared.
From the present study, it can be concluded that the automated hematology analyzer readings correlated well with readings by the standard manual method, although the latter method gave additional diagnostic information on the blood pictures. While patients' care and laboratory operations could be optimized by using manual microscopic examination as a reflective substitute for automated methods, usage of automated method would ease our workload and save time for patients.</description><subject>Blood</subject><subject>Blood diseases</subject><subject>Demographic aspects</subject><subject>Diagnosis</subject><subject>Medical examination</subject><subject>Methods</subject><subject>Parameter estimation</subject><subject>Research article</subject><issn>1472-6890</issn><issn>1472-6890</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><recordid>eNp1kk2LFDEQhoMo7rp69SgBD556zdd00hdhGPxYXPSggreQTlfPRDrJmHSL7cHfbtpZx11cyaGKqrce3hSF0GNKzilV9XMqJKtq1ZCKkorfQafHwt1r-Ql6kPMXQqhUlN1HJ4wIWpNanaKfm-j3JrkcA4493hnwZoxD3DprBlw6xsMIKeNuCd4F6HA743EH-MOcPXzHbz_jCrOLd9hMYyzDRfCXMmMTzDD_gFSS7veYN2EqaBunMOaH6F5vhgyPruIZ-vTq5cfNm-ry_euLzfqyaldUjJVsGy6VJEoxYKQRlIve1g0RDNqmsRY6Rluwsparpu0oh46WbwtB5IqImlJ-hl4cuPup9dBZCGMyg94n502adTRO3-wEt9Pb-E0zJbkQogDWB0Dr4n8ANzs2er3sXy_715RoXhjPrkyk-HWCPGrvsoVhMAHilLXkfCWkkKoonx6UWzOAdqGPhWkXtV4zVivelA0U1fktqvI68M7GAL0r9dsGbIo5J-iP_ou_5Zz-dfzk-tqO8j_3w38B8TnGBA</recordid><startdate>20100423</startdate><enddate>20100423</enddate><creator>Ike, Samuel O</creator><creator>Nubila, Thomas</creator><creator>Ukaejiofo, Ernest O</creator><creator>Nubila, Imelda N</creator><creator>Shu, Elvis N</creator><creator>Ezema, Ifeyinwa</creator><general>BioMed Central Ltd</general><general>BioMed Central</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>20100423</creationdate><title>Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts</title><author>Ike, Samuel O ; Nubila, Thomas ; Ukaejiofo, Ernest O ; Nubila, Imelda N ; Shu, Elvis N ; Ezema, Ifeyinwa</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Blood</topic><topic>Blood diseases</topic><topic>Demographic aspects</topic><topic>Diagnosis</topic><topic>Medical examination</topic><topic>Methods</topic><topic>Parameter estimation</topic><topic>Research article</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Ike, Samuel O</creatorcontrib><creatorcontrib>Nubila, Thomas</creatorcontrib><creatorcontrib>Ukaejiofo, Ernest O</creatorcontrib><creatorcontrib>Nubila, Imelda N</creatorcontrib><creatorcontrib>Shu, Elvis N</creatorcontrib><creatorcontrib>Ezema, Ifeyinwa</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>BMC clinical pathology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ike, Samuel O</au><au>Nubila, Thomas</au><au>Ukaejiofo, Ernest O</au><au>Nubila, Imelda N</au><au>Shu, Elvis N</au><au>Ezema, Ifeyinwa</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts</atitle><jtitle>BMC clinical pathology</jtitle><addtitle>BMC Clin Pathol</addtitle><date>2010-04-23</date><risdate>2010</risdate><volume>10</volume><issue>1</issue><spage>3</spage><epage>3</epage><pages>3-3</pages><artnum>3</artnum><issn>1472-6890</issn><eissn>1472-6890</eissn><abstract>This study was designed to determine the correlation between heamatological parameters by Sysmex KX-21N automated hematology analyzer with the manual methods.
Sixty (60) subjects were randomly selected from both apparently healthy subjects and those who have different blood disorders from the University of Teaching Hospital (UNTH), Ituku-Ozalla, Enugu, Enugu State, Nigeria. Three (3)mls of venous blood sample was collected aseptically from each subject into tri-potassium ethylenediamine tetra-acetic acid (K3EDTA) for the analysis of haematological parameters using the automated and the manual methods.
The blood film report by the manual method showed that 50% of the subjects were normocytic-normochromic while the other 50% revealed different abnormal blood pictures. Also, there were statistically significant differences (p < 0.05) in mean cell hemoglobin concentrations (MCHC) between the two methods. Similarly, the mean (S.E) values of hemoglobin, packed cell volume, platelet and total white cell counts demonstrated statistically significant difference (p < 0.001) and correlated positively when both methods were compared.
From the present study, it can be concluded that the automated hematology analyzer readings correlated well with readings by the standard manual method, although the latter method gave additional diagnostic information on the blood pictures. While patients' care and laboratory operations could be optimized by using manual microscopic examination as a reflective substitute for automated methods, usage of automated method would ease our workload and save time for patients.</abstract><cop>England</cop><pub>BioMed Central Ltd</pub><pmid>20416068</pmid><doi>10.1186/1472-6890-10-3</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1472-6890 |
ispartof | BMC clinical pathology, 2010-04, Vol.10 (1), p.3-3, Article 3 |
issn | 1472-6890 1472-6890 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_2873444 |
source | ProQuest - Publicly Available Content Database; PubMed Central |
subjects | Blood Blood diseases Demographic aspects Diagnosis Medical examination Methods Parameter estimation Research article |
title | Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T11%3A46%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Comparison%20of%20haematological%20parameters%20determined%20by%20the%20Sysmex%20KX%20-%202IN%20automated%20haematology%20analyzer%20and%20the%20manual%20counts&rft.jtitle=BMC%20clinical%20pathology&rft.au=Ike,%20Samuel%20O&rft.date=2010-04-23&rft.volume=10&rft.issue=1&rft.spage=3&rft.epage=3&rft.pages=3-3&rft.artnum=3&rft.issn=1472-6890&rft.eissn=1472-6890&rft_id=info:doi/10.1186/1472-6890-10-3&rft_dat=%3Cgale_pubme%3EA226839134%3C/gale_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-b514t-7b937870882e2094134fc69042eb99cced21bec76759bd13ed168944075046113%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=733547478&rft_id=info:pmid/20416068&rft_galeid=A226839134&rfr_iscdi=true |