Loading…

Photon counting, censor corrections, and lifetime imaging for improved detection in two-photon microscopy

We present a high-speed photon counter for use with two-photon microscopy. Counting pulses of photocurrent, as opposed to analog integration, maximizes the signal-to-noise ratio so long as the uncertainty in the count does not exceed the gain-noise of the photodetector. Our system extends this impro...

Full description

Saved in:
Bibliographic Details
Published in:Journal of neurophysiology 2011-06, Vol.105 (6), p.3106-3113
Main Authors: Driscoll, Jonathan D, Shih, Andy Y, Iyengar, Satish, Field, Jeffrey J, White, G Allen, Squier, Jeffrey A, Cauwenberghs, Gert, Kleinfeld, David
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We present a high-speed photon counter for use with two-photon microscopy. Counting pulses of photocurrent, as opposed to analog integration, maximizes the signal-to-noise ratio so long as the uncertainty in the count does not exceed the gain-noise of the photodetector. Our system extends this improvement through an estimate of the count that corrects for the censored period after detection of an emission event. The same system can be rapidly reconfigured in software for fluorescence lifetime imaging, which we illustrate by distinguishing between two spectrally similar fluorophores in an in vivo model of microstroke.