Loading…
Morphological evaluation of eccentric sets guide-plates of dcp-l 4.5 mm
To carry out isolated and comparative evaluations of the measurements of the set eccentric guide plates used in 4.5mm surgical implants, and to determine the effect of these measurements on compression strength. Four eccentric guides, four large dynamic compression plates (L-DCP) from four local man...
Saved in:
Published in: | Acta ortopedica brasileira 2012, Vol.20 (1), p.13-16 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To carry out isolated and comparative evaluations of the measurements of the set eccentric guide plates used in 4.5mm surgical implants, and to determine the effect of these measurements on compression strength.
Four eccentric guides, four large dynamic compression plates (L-DCP) from four local manufacturers, and a Vonder(®) 200 mm caliper brand were used. Five standard parameter measurements were created for the set eccentric guide-plate, which were identified as A to E. Four sets were made, using materials of the same factory, and identified as groups I to IV. The analyses were performed by measuring all the parameters from a ventral view of the plate, with the eccentric guide placed in the plate hole.
Groups I and II showed the same values for all the parameters. All the groups showed the same measurements for E = 8.15 e B = 3.60. Group III: A = 8.10mm, C = 3.25mm, D = 1.25mm. Group IV: A = 7.00mm, C = 3.10mm, D = 0.30mm. Maximum compression force was (F Max.): Group I 80.58 N, Group. II: F Max. 81.63 N, Group. III: F Max. 36.32N, Group. IV: F Max. 37.52N.
The measurements evaluated show a lack of standardization in the manufacture of orthopedic instruments and its effects on the values for compression strength.
Level III, analytical study. |
---|---|
ISSN: | 1413-7852 1809-4406 |
DOI: | 10.1590/S1413-78522012000100002 |