Loading…

Visualisation of the Bonebridge by means of CT and CBCT

With the Bonebridge, a new bone-anchored hearing aid has been available since March 2012. The objective of the study was to analyse the visualisation of the implant itself as well as its impact on the representation of the bony structures of the petrosal bone in CT, MRI and cone beam CT (CBCT). The...

Full description

Saved in:
Bibliographic Details
Published in:European journal of medical research 2013-09, Vol.18 (1), p.30-30, Article 30
Main Authors: Güldner, Christian, Heinrichs, Julia, Weiß, Rainer, Zimmermann, Annette Paula, Dassinger, Benjamin, Bien, Siegfried, Werner, Jochen Alfred, Diogo, Isabell
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853
cites cdi_FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853
container_end_page 30
container_issue 1
container_start_page 30
container_title European journal of medical research
container_volume 18
creator Güldner, Christian
Heinrichs, Julia
Weiß, Rainer
Zimmermann, Annette Paula
Dassinger, Benjamin
Bien, Siegfried
Werner, Jochen Alfred
Diogo, Isabell
description With the Bonebridge, a new bone-anchored hearing aid has been available since March 2012. The objective of the study was to analyse the visualisation of the implant itself as well as its impact on the representation of the bony structures of the petrosal bone in CT, MRI and cone beam CT (CBCT). The Bonebridge was implanted unilaterally in two completely prepared human heads. The radiological imaging by means of CBCT, 64-slice CT, 1.5-T and 3.0-T MRI was conducted both preoperatively and postoperatively. The images were subsequently evaluated from both the ENT medical and nd radiological perspectives. As anticipated, no visualisation of the implant or of the petrosal bones could be realised on MRI because of the interactive technology and the magnet artefact. In contrast, an excellent evaluability of the implant itself as well as of the surrounding neurovascular structures (sinus sigmoideus, skull base, middle ear, inner ear, inner auditory canal) was exhibited in both the CT and in the CBCT. The Bonebridge can be excellently imaged with the radiological imaging technologies of CT and CBCT. In the process, CBCT shows discrete advantages in comparison with CT. No relevant restrictions in image quality in the evaluation of the bony structures of the petrosal bones could be seen.
doi_str_mv 10.1186/2047-783X-18-30
format article
fullrecord <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_3844407</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A534829577</galeid><sourcerecordid>A534829577</sourcerecordid><originalsourceid>FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853</originalsourceid><addsrcrecordid>eNp1kt9L5DAQx8OhqKjPvh0FQXypJk3SJC-CW84fIPiyd9xbmHanu5G20aYV_O8vZb1lFzR5SJj5zJeZb0LIGaNXjOn8OqNCpUrzvynTKac_yNEmsrd1PySnIbzQuPIsV8YckMNMUCoM5UdE_XFhhMYFGJzvEl8nwwqTme-w7N1iiUn5kbQIXZhSxTyBbpEUs2J-QvZraAKefp7H5Pfdr3nxkD493z8Wt09pKSUdUlA5h1LUSjCOtal1TjOIPWjMcgGsRA0VSGRMKoMGdclYlikGFcqccy35MblZ676OZYuLCruhh8a-9q6F_sN6cHY307mVXfp3y7UQgqooMFsLlM5_I7CbqXxrJ-_s5J1l2nIaRS4_u-j924hhsK0LFTYNdOjHYJmQhknDjYno-RpdQoPWdbWPqtWE21vJhc6MVFNXV19QcS-wdVV0v3YxvlNwsVWwQmiGVfDNOL1a2AWv12DV-xB6rDejMmqnb_PFcD-3Ld7w_z8J_weWwrpP</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>1459159399</pqid></control><display><type>article</type><title>Visualisation of the Bonebridge by means of CT and CBCT</title><source>PubMed Central</source><creator>Güldner, Christian ; Heinrichs, Julia ; Weiß, Rainer ; Zimmermann, Annette Paula ; Dassinger, Benjamin ; Bien, Siegfried ; Werner, Jochen Alfred ; Diogo, Isabell</creator><creatorcontrib>Güldner, Christian ; Heinrichs, Julia ; Weiß, Rainer ; Zimmermann, Annette Paula ; Dassinger, Benjamin ; Bien, Siegfried ; Werner, Jochen Alfred ; Diogo, Isabell</creatorcontrib><description>With the Bonebridge, a new bone-anchored hearing aid has been available since March 2012. The objective of the study was to analyse the visualisation of the implant itself as well as its impact on the representation of the bony structures of the petrosal bone in CT, MRI and cone beam CT (CBCT). The Bonebridge was implanted unilaterally in two completely prepared human heads. The radiological imaging by means of CBCT, 64-slice CT, 1.5-T and 3.0-T MRI was conducted both preoperatively and postoperatively. The images were subsequently evaluated from both the ENT medical and nd radiological perspectives. As anticipated, no visualisation of the implant or of the petrosal bones could be realised on MRI because of the interactive technology and the magnet artefact. In contrast, an excellent evaluability of the implant itself as well as of the surrounding neurovascular structures (sinus sigmoideus, skull base, middle ear, inner ear, inner auditory canal) was exhibited in both the CT and in the CBCT. The Bonebridge can be excellently imaged with the radiological imaging technologies of CT and CBCT. In the process, CBCT shows discrete advantages in comparison with CT. No relevant restrictions in image quality in the evaluation of the bony structures of the petrosal bones could be seen.</description><identifier>ISSN: 2047-783X</identifier><identifier>ISSN: 0949-2321</identifier><identifier>EISSN: 2047-783X</identifier><identifier>DOI: 10.1186/2047-783X-18-30</identifier><identifier>PMID: 24004903</identifier><language>eng</language><publisher>England: BioMed Central Ltd</publisher><subject>Bones ; Cone-Beam Computed Tomography ; CT imaging ; Hearing Aids ; Humans ; Imaging systems ; Imaging, Three-Dimensional ; Intraoperative Care ; Magnetic resonance imaging ; Physiological aspects ; Prostheses and Implants</subject><ispartof>European journal of medical research, 2013-09, Vol.18 (1), p.30-30, Article 30</ispartof><rights>COPYRIGHT 2013 BioMed Central Ltd.</rights><rights>Copyright © 2013 Güldner et al.; licensee BioMed Central Ltd. 2013 Güldner et al.; licensee BioMed Central Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853</citedby><cites>FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC3844407/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC3844407/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,881,27901,27902,53766,53768</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/24004903$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Güldner, Christian</creatorcontrib><creatorcontrib>Heinrichs, Julia</creatorcontrib><creatorcontrib>Weiß, Rainer</creatorcontrib><creatorcontrib>Zimmermann, Annette Paula</creatorcontrib><creatorcontrib>Dassinger, Benjamin</creatorcontrib><creatorcontrib>Bien, Siegfried</creatorcontrib><creatorcontrib>Werner, Jochen Alfred</creatorcontrib><creatorcontrib>Diogo, Isabell</creatorcontrib><title>Visualisation of the Bonebridge by means of CT and CBCT</title><title>European journal of medical research</title><addtitle>Eur J Med Res</addtitle><description>With the Bonebridge, a new bone-anchored hearing aid has been available since March 2012. The objective of the study was to analyse the visualisation of the implant itself as well as its impact on the representation of the bony structures of the petrosal bone in CT, MRI and cone beam CT (CBCT). The Bonebridge was implanted unilaterally in two completely prepared human heads. The radiological imaging by means of CBCT, 64-slice CT, 1.5-T and 3.0-T MRI was conducted both preoperatively and postoperatively. The images were subsequently evaluated from both the ENT medical and nd radiological perspectives. As anticipated, no visualisation of the implant or of the petrosal bones could be realised on MRI because of the interactive technology and the magnet artefact. In contrast, an excellent evaluability of the implant itself as well as of the surrounding neurovascular structures (sinus sigmoideus, skull base, middle ear, inner ear, inner auditory canal) was exhibited in both the CT and in the CBCT. The Bonebridge can be excellently imaged with the radiological imaging technologies of CT and CBCT. In the process, CBCT shows discrete advantages in comparison with CT. No relevant restrictions in image quality in the evaluation of the bony structures of the petrosal bones could be seen.</description><subject>Bones</subject><subject>Cone-Beam Computed Tomography</subject><subject>CT imaging</subject><subject>Hearing Aids</subject><subject>Humans</subject><subject>Imaging systems</subject><subject>Imaging, Three-Dimensional</subject><subject>Intraoperative Care</subject><subject>Magnetic resonance imaging</subject><subject>Physiological aspects</subject><subject>Prostheses and Implants</subject><issn>2047-783X</issn><issn>0949-2321</issn><issn>2047-783X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2013</creationdate><recordtype>article</recordtype><recordid>eNp1kt9L5DAQx8OhqKjPvh0FQXypJk3SJC-CW84fIPiyd9xbmHanu5G20aYV_O8vZb1lFzR5SJj5zJeZb0LIGaNXjOn8OqNCpUrzvynTKac_yNEmsrd1PySnIbzQuPIsV8YckMNMUCoM5UdE_XFhhMYFGJzvEl8nwwqTme-w7N1iiUn5kbQIXZhSxTyBbpEUs2J-QvZraAKefp7H5Pfdr3nxkD493z8Wt09pKSUdUlA5h1LUSjCOtal1TjOIPWjMcgGsRA0VSGRMKoMGdclYlikGFcqccy35MblZ676OZYuLCruhh8a-9q6F_sN6cHY307mVXfp3y7UQgqooMFsLlM5_I7CbqXxrJ-_s5J1l2nIaRS4_u-j924hhsK0LFTYNdOjHYJmQhknDjYno-RpdQoPWdbWPqtWE21vJhc6MVFNXV19QcS-wdVV0v3YxvlNwsVWwQmiGVfDNOL1a2AWv12DV-xB6rDejMmqnb_PFcD-3Ld7w_z8J_weWwrpP</recordid><startdate>20130903</startdate><enddate>20130903</enddate><creator>Güldner, Christian</creator><creator>Heinrichs, Julia</creator><creator>Weiß, Rainer</creator><creator>Zimmermann, Annette Paula</creator><creator>Dassinger, Benjamin</creator><creator>Bien, Siegfried</creator><creator>Werner, Jochen Alfred</creator><creator>Diogo, Isabell</creator><general>BioMed Central Ltd</general><general>BioMed Central</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>20130903</creationdate><title>Visualisation of the Bonebridge by means of CT and CBCT</title><author>Güldner, Christian ; Heinrichs, Julia ; Weiß, Rainer ; Zimmermann, Annette Paula ; Dassinger, Benjamin ; Bien, Siegfried ; Werner, Jochen Alfred ; Diogo, Isabell</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2013</creationdate><topic>Bones</topic><topic>Cone-Beam Computed Tomography</topic><topic>CT imaging</topic><topic>Hearing Aids</topic><topic>Humans</topic><topic>Imaging systems</topic><topic>Imaging, Three-Dimensional</topic><topic>Intraoperative Care</topic><topic>Magnetic resonance imaging</topic><topic>Physiological aspects</topic><topic>Prostheses and Implants</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Güldner, Christian</creatorcontrib><creatorcontrib>Heinrichs, Julia</creatorcontrib><creatorcontrib>Weiß, Rainer</creatorcontrib><creatorcontrib>Zimmermann, Annette Paula</creatorcontrib><creatorcontrib>Dassinger, Benjamin</creatorcontrib><creatorcontrib>Bien, Siegfried</creatorcontrib><creatorcontrib>Werner, Jochen Alfred</creatorcontrib><creatorcontrib>Diogo, Isabell</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>European journal of medical research</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Güldner, Christian</au><au>Heinrichs, Julia</au><au>Weiß, Rainer</au><au>Zimmermann, Annette Paula</au><au>Dassinger, Benjamin</au><au>Bien, Siegfried</au><au>Werner, Jochen Alfred</au><au>Diogo, Isabell</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Visualisation of the Bonebridge by means of CT and CBCT</atitle><jtitle>European journal of medical research</jtitle><addtitle>Eur J Med Res</addtitle><date>2013-09-03</date><risdate>2013</risdate><volume>18</volume><issue>1</issue><spage>30</spage><epage>30</epage><pages>30-30</pages><artnum>30</artnum><issn>2047-783X</issn><issn>0949-2321</issn><eissn>2047-783X</eissn><abstract>With the Bonebridge, a new bone-anchored hearing aid has been available since March 2012. The objective of the study was to analyse the visualisation of the implant itself as well as its impact on the representation of the bony structures of the petrosal bone in CT, MRI and cone beam CT (CBCT). The Bonebridge was implanted unilaterally in two completely prepared human heads. The radiological imaging by means of CBCT, 64-slice CT, 1.5-T and 3.0-T MRI was conducted both preoperatively and postoperatively. The images were subsequently evaluated from both the ENT medical and nd radiological perspectives. As anticipated, no visualisation of the implant or of the petrosal bones could be realised on MRI because of the interactive technology and the magnet artefact. In contrast, an excellent evaluability of the implant itself as well as of the surrounding neurovascular structures (sinus sigmoideus, skull base, middle ear, inner ear, inner auditory canal) was exhibited in both the CT and in the CBCT. The Bonebridge can be excellently imaged with the radiological imaging technologies of CT and CBCT. In the process, CBCT shows discrete advantages in comparison with CT. No relevant restrictions in image quality in the evaluation of the bony structures of the petrosal bones could be seen.</abstract><cop>England</cop><pub>BioMed Central Ltd</pub><pmid>24004903</pmid><doi>10.1186/2047-783X-18-30</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2047-783X
ispartof European journal of medical research, 2013-09, Vol.18 (1), p.30-30, Article 30
issn 2047-783X
0949-2321
2047-783X
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_3844407
source PubMed Central
subjects Bones
Cone-Beam Computed Tomography
CT imaging
Hearing Aids
Humans
Imaging systems
Imaging, Three-Dimensional
Intraoperative Care
Magnetic resonance imaging
Physiological aspects
Prostheses and Implants
title Visualisation of the Bonebridge by means of CT and CBCT
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-01T04%3A03%3A04IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Visualisation%20of%20the%20Bonebridge%20by%20means%20of%20CT%20and%20CBCT&rft.jtitle=European%20journal%20of%20medical%20research&rft.au=G%C3%BCldner,%20Christian&rft.date=2013-09-03&rft.volume=18&rft.issue=1&rft.spage=30&rft.epage=30&rft.pages=30-30&rft.artnum=30&rft.issn=2047-783X&rft.eissn=2047-783X&rft_id=info:doi/10.1186/2047-783X-18-30&rft_dat=%3Cgale_pubme%3EA534829577%3C/gale_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-b550t-a763ab4f7413ef9f8602a7998e264a1be8aca5e11579e9e8b112271ace5633853%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=1459159399&rft_id=info:pmid/24004903&rft_galeid=A534829577&rfr_iscdi=true