Study on Construction of a Medical X-Ray Direct Digital Radiography System and Hybrid Preprocessing Methods

We construct a medical X-ray direct digital radiography (DDR) system based on a CCD (charge-coupled devices) camera. For the original images captured from X-ray exposure, computer first executes image flat-field correction and image gamma correction, and then carries out image contrast enhancement....

Full description

Saved in:
Bibliographic Details
Published in:Computational and mathematical methods in medicine 2014-01, Vol.2014 (2014), p.1-7
Main Authors: Ren, Yong, Wu, Sheng, Wang, Mijian, Cen, Zhongjie
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We construct a medical X-ray direct digital radiography (DDR) system based on a CCD (charge-coupled devices) camera. For the original images captured from X-ray exposure, computer first executes image flat-field correction and image gamma correction, and then carries out image contrast enhancement. A hybrid image contrast enhancement algorithm which is based on sharp frequency localization-contourlet transform (SFL-CT) and contrast limited adaptive histogram equalization (CLAHE), is proposed and verified by the clinical DDR images. Experimental results show that, for the medical X-ray DDR images, the proposed comprehensive preprocessing algorithm can not only greatly enhance the contrast and detail information, but also improve the resolution capability of DDR system.