Simultaneous quantitation of 17α-hydroxyprogesterone caproate, 17α-hydroxyprogesterone and progesterone in human plasma using high-performance liquid chromatography–mass spectrometry (HPLC–MS/MS)
A sensitive and specific assay method for the simultaneous quantitation of 17α-hydroxyprogesterone caproate (17-OHPC), 17α-hydroxyprogesterone (17-OHP), and progesterone (P) in human plasma using high-performance liquid chromatography and tandem mass spectrometry (LC–MS/MS) was developed and validat...
Saved in:
Published in: | Journal of pharmaceutical and biomedical analysis 2008-12, Vol.48 (4), p.1174-1180 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A sensitive and specific assay method for the simultaneous quantitation of 17α-hydroxyprogesterone caproate (17-OHPC), 17α-hydroxyprogesterone (17-OHP), and progesterone (P) in human plasma using high-performance liquid chromatography and tandem mass spectrometry (LC–MS/MS) was developed and validated. Plasma samples were processed by a solid phase extraction (SPE) procedure using Oasis
® HLB extraction cartridge prior to chromatography. Medroxyprogestrone acetate (MPA) was used as the internal standard. The compounds were separated using Waters C18 Symmetry analytical column (3.5
μm, 2.1
mm
×
50
mm) using a gradient elusion with a mobile phase consisting of 5% methanol in water [A] and methanol [B], with ammonium acetate (2
mM) and formic acid (0.1%) being added to both [A] and [B], at a flow rate 0.3
ml/min. The retention times for 17-OHPC, 17-OHP, P and MPA were 4.5, 1.5, 2.5 and 2.2
min, respectively, with a total run time of 7
min. The analytes were detected by a Micromass Quattro Micro triple quadrupole mass spectrometer in positive electron spray ionization (ESI) mode using multiple reaction monitoring (MRM). The extracted ions monitored following MRM transitions were
m/
z 429.10
→
313.10 for 17-OHPC,
m/
z 331.17
→
97.00 for 17-OHP,
m/
z 315.15
→
109.00 for P and
m/
z 387.15
→
327.25 for MPA (IS). The assay was linear over the range 1–200
ng/ml for 17-OHPC and 17-OHP, and 2–400
ng/ml for P, when 0.4
ml of plasma was used in the extraction. The overall intra- and inter-day assay variation was |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/j.jpba.2008.08.024 |