Fatal outcome arising from use of a sutureless "corkscrew" epicardial pacing electrode inserted into apex of left ventricle
A 59-year-old man is described in whom the insertion of an epicardial sutureless "corkscrew" electrode resulted in fatal ventricular perforation. Fatal myocardial perforation can occur with this electrode and the apex of the left ventricle should never be used as the site of insertion. Nec...
Saved in:
Published in: | British Heart Journal 1976-12, Vol.38 (12), p.1359-1362 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A 59-year-old man is described in whom the insertion of an epicardial sutureless "corkscrew" electrode resulted in fatal ventricular perforation. Fatal myocardial perforation can occur with this electrode and the apex of the left ventricle should never be used as the site of insertion. Necropsy also showed that the transvenous right ventricular electrode, inserted one year previously, had penetrated a tricuspid leaflet. This could have accounted for the ensuing pacing failure. |
---|---|