A Summary of Heat-Flux Sensor Calibration Data
This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(-2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the...
Saved in:
Published in: | Journal of research of the National Institute of Standards and Technology 2005-03, Vol.110 (2), p.97-100 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(-2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the NIST and the customer measured responsivity values showed that the measurements agree within 3 % for more than half the number of sensors tested, so far. Considering the variation in the customer calibration techniques and the wide measuring range of the sensors used in the calibration, the agreement is encouraging. |
---|---|
ISSN: | 1044-677X 2165-7254 |
DOI: | 10.6028/jres.110.009 |