Loading…

A Summary of Heat-Flux Sensor Calibration Data

This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(-2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the...

Full description

Saved in:
Bibliographic Details
Published in:Journal of research of the National Institute of Standards and Technology 2005-03, Vol.110 (2), p.97-100
Main Authors: Murthy, A V, Fraser, G T, DeWitt, D P
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(-2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the NIST and the customer measured responsivity values showed that the measurements agree within 3 % for more than half the number of sensors tested, so far. Considering the variation in the customer calibration techniques and the wide measuring range of the sensors used in the calibration, the agreement is encouraging.
ISSN:1044-677X
2165-7254
DOI:10.6028/jres.110.009