Inaccuracy of Urine Output Measurements due to Urinary Retention in Catheterized Patients in the Burn ICU

Electronic urinary output monitors, intended to provide urine output information to guide fluid therapy during burn resuscitation, can be inaccurate because of airlocks causing urine retention in the drainage tube and bladder. In this study, the authors explore the effects of airlock formation on ur...

Full description

Saved in:
Bibliographic Details
Published in:Journal of burn care & research 2017-01, Vol.38 (1), p.e409-e417
Main Authors: Kramer, George C, Luxon, Evan, Wolf, Jordan, Burnett, Daniel R, Nanduri, Devyani, Friedman, Bruce C
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Electronic urinary output monitors, intended to provide urine output information to guide fluid therapy during burn resuscitation, can be inaccurate because of airlocks causing urine retention in the drainage tube and bladder. In this study, the authors explore the effects of airlock formation on urine output measured using an electronic urinary output monitor connected to either a standard commercial drainage tubing system or a drainage tubing system with an automated airlock clearing mechanism. In a multicenter study in the burn intensive care unit, urine output was compared between 10 control patients with a standard commercial drainage tubing system and 10 test patients with a novel automated airlock clearing drainage tubing system. The comparison was focused on identifying the number and magnitude of surges in urinary output because of airlocks and associated periods of false oliguria. In the control group, 5 of 10 (50%) patients had drainage line flow impediments from 8 airlocks. In addition, control patients experienced six associated periods of false oliguria. Airlock surge volumes ranged from 50 to 329 ml, and false oliguria duration ranged from 39.4 to 185.2 minutes. In the test group, 0 of 10 (0%) patients had drainage line impediments from airlocks (P < .01), and hence, there were no periods of false oliguria. Airlocks and associated periods of false oliguria occur with standard commercial drainage tubing and are eliminated using an automated airlock clearing drainage tube. Electronic urinary output monitoring with self-clearing drainage has the potential to improve tracking of real-time urine output and decrease caregiver workload.
ISSN:1559-047X
1559-0488
DOI:10.1097/BCR.0000000000000405