The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube

Although serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new...

Full description

Saved in:
Bibliographic Details
Published in:Biochemia Medica 2017-10, Vol.27 (3), p.030706-030706
Main Authors: Arslan, Fatma Demet, Karakoyun, Inanc, Basok, Banu Isbilen, Aksit, Merve Zeytinli, Baysoy, Anil, Ozturk, Yasemin Kilic, Guclu, Yusuf Adnan, Duman, Can
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Although serum-providing blood tubes with a barrier are still widely used due to their significant advantages, the use of blood tubes with a barrier to provide plasma is becoming widespread. We compared 22 analytes in a BD Vacutainer® Barricor LH Plasma tube for local clinical validation of this new lithium heparin tube with a barrier. Samples from 44 volunteers were collected in different tubes (Becton Dickinson and Company): Z tube without additive (reference), clot-activator tube with gel (SST), lithium heparin tube without gel (LiH), and lithium heparin tube with barrier (Barricor). Analyte concentrations in different tubes were compared with the reference tube. All tubes were also evaluated according to additional testing (different centrifugation durations, blood-sampling techniques and individual differences). Aspartate aminotransferase (AST), glucose (Glc), potassium (K), lactate dehydrogenase (LD), sodium (Na), and total protein (TP) had a significant bias in Barricor (9.19%, - 3.24%, - 4.88%, 21.60%, - 0.40%, 5.03%, respectively) relative to the reference tube. There was no statistical difference between different centrifugation durations and individual differences for AST, K and LD in LiH and/or Barricor (P > 0.05). There was a significant bias for LD between LiH and Barricor in terms of blood-sampling techniques (21.2% and 12.4%, respectively). Recently, the use of plasma has become prominent due to some of its advantages. In this study, plasma AST, K, LD, Glc and TP levels in Barricor were clinically different in comparison to serum. The results of additional tests showed that higher levels of LD in Barricor did not result from haemolysis, and they might be related to other factors including number of platelets, cellular fragility, or functional environment.
ISSN:1330-0962
1846-7482
DOI:10.11613/BM.2017.030706